Installation Guide1-800-257-3872 Toll Free1-978-264-0292 Faxwww.setra.com Web SiteModel 224Ultra-High Purity Flow-ThroughPressure TransducerISO✓Certif
2Setra Model 224Ultra High Purity Flow-Through Pressure Transducer1.0 GENERAL INFORMATIONEvery Model 224 has been tested and calibrated before shipmen
2.4 VentingModel 224 transducers are vented through the electronics housing.3.0 ELECTRICAL INSTALLATION3.1 Voltage Output UnitsThe Model 224 voltage
44.0 CALIBRATIONThe 224 transducer is factory calibrated to the specific input pressure range vs.output voltage or current and should require no fie
4.1 Voltage Output Zero AdjustmentWhile monitoring the voltage between the positive output (+OUT) and common(COM), and with the pressure port open to
64.2 Voltage Output Span Adjustment(Complete the zero adjustment before setting span.)Span or full scale output adjustments should only be performed
6.0 RETURNING PRODUCTS FOR REPAIRPlease contact a Setra application engineer (800-257-3872, 978-263-1400) beforereturning unit for repair to review in