Specifications
Pressure media: Clean dry nonconductive noncorrosive gas
Accuracy:
accuracy
1
±0.02% full scale at 70°F (21°C)
nonlinearity ±0.012% full scale (End point)
hysteresis 0.010% full scale
nonrepeatability ±0.010% full scale
thermal zero shift
2
±0.002% FS/°F (0.004% FS/°C)
thermal span shift
2
±0.001% FS/°F (0.002% FS/°C)
altitude resolution 1 ft. (4 ft. for 100 psi range)
stability, 24 hours ±0.005% full scale
stability, 30 days ±0.02% full scale
stability, 1 year ±0.05% full scale
Output: Bidirectional RS-232 interface
Physical:
pressure fitting 1/8” barbed male fitting
pressure port 10-32 internal thread
height 4.0”
width 3.5”
depth 5.25”
weight 2.4 pounds
Power: 5 VDC ±1%, 70 mA; 17 mA w/200
microampere sleep mode option.
Notes:
1. Accuracy as RSS of non-linearity, hysteresis, and non-repeatability.
2. Unit calibrated at 70°F. Maximum thermal error is computed from this datum.