Setra-system Model 270 Manual de usuario Pagina 3

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 5
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 2
INSTALLATION INSTRUCTIONS FOR MULTIPLE HOOK-UP
OF 4-WIRE SETRA PRESSURE TRANSDUCERS
OPTION #1
Uses a single, ground referenced power supply for excitation and separate, isolated (not ground refer-
enced) readout or “differential input” to a data acquisition system for each pressure transducer output.
OPTION #2
Uses a separate, isolated power supply for each pressure transducer’s excitation and a single, ground
referenced readout or “single ended input” to a data acquisition system for all of the outputs.
Power Supply (Nominal 24 VDC)
PT - Pressure Transducer (4 Wire Circuit)
DVM/DAS - Digital Volt Meter or Data Acquisition System
Vista de pagina 2
1 2 3 4 5

Comentarios a estos manuales

Sin comentarios